Coriolis Mass Flow Meter Lab Report
Abstract This is an experiment about procedures of calibration of a Coriolis mass flow meter. To do the experiment, water is pumped from a reservoir and its flow rate is set by a tap, the height drop is examined by eye and the time is recorded for a period during the flow. This procedure is arranged for the calibration of the Coriolis meter by comparing the calculated mass flow and the read value. Although, it is a reasonable aspect for the calibration which is comparing digital and analog measurement methods, the control mechanism of the experiment, in this case human eye and a ruler placed on the transparent reservoir could give erroneous results. In the report, it is also discussed how this kind of errors could be fixed and what solutions …show more content…
Each coriolis flow meter contains a tube and an exciter that makes this tube to oscillate constantly. These parts can be seen from the figure 3. Figure 3: Different aspects of a coriolis mass flow meter When there is no flow in the pipe the measuring tube oscillates uniformly. In addition, sensors are located at the inlet and the outlet of the tube. These sensors have ability to notice this basic oscillation perfectly when the fluid starts to flow in the measuring tube. Additional twisting is applied on the oscillation because of the liquids inertia. As a result of Coriolis Effect, the inlet and the outlet parts of the tube oscillate in different directions simultaneously. These very sensitive sensors gather this change in the tube oscillation in terms of time and space that is called the phase shift. How much liquid or gas is flowing through the pipe can be measured using this phase shift. The higher flow velocity and therefore the total flow, the greater the deflection of the oscillating measuring tube. A typical coriolis mass flow meter can measure not only mass flow but also volume flow, density, temperature, and viscosity with the help of these extremely sensitive …show more content…
Assuming that C_u has negligible effect on the equation and knowing that Ω is the angular velocity of the vibrating tube that can be shown using following equation:
Ω=〖Ω 〗_(0 ) cos⁡ωt (5)
Equation (4) can be solved using information above and this particular solution is obtained: θ=θ_0 cos⁡ωt = (2K〖〖Ω 〗_(0 ) Q〗_m dl)/(K_u-I_u ω^2 ) cos⁡ωt (6)
Moreover, there is a time lag between two sides of the tube and it can be found from velocity over displacement such that τ=(θ_0 d)/(Ω_0 l)=(2KQ_m d^2)/(K_u-I_u ω^2 ) (7)
Then, the equation for mass flow rate can be obtained after measuring the time lag and the equation is :
Q_m=(K_u-I_u ω^2)/(2Kd^2 ) τ (8)
Experimental Setup &

You May Also Find These Documents Helpful

  • Satisfactory Essays

    AP Physics Slinky Lab

    • 468 Words
    • 4 Pages

    m) and it was moved side to side to create a longitudinal wave and the period was…

    • 468 Words
    • 4 Pages
    Satisfactory Essays
  • Satisfactory Essays

    Priple Beam Balance

    • 213 Words
    • 1 Page

    A graduated cylinder was used for measuring the volumes of the liquid. When measuring the volumes be sure to be eye level with the meniscus. Any other position when looksing at the container will cause an error in the experiment.…

    • 213 Words
    • 1 Page
    Satisfactory Essays
  • Good Essays

    Basic Lab Measurements

    • 465 Words
    • 2 Pages

    This lab experiment displays the accuracy of scientific measuring equipment. This experiment also helps to familiarize the students with these instruments.…

    • 465 Words
    • 2 Pages
    Good Essays
  • Satisfactory Essays

    Chem Lab 2 Report

    • 942 Words
    • 9 Pages

    Procedure: In the experiment, I use devices such as a ruler, beakers, flasks, a scale and cylinders to measure the values of each object and liquid.…

    • 942 Words
    • 9 Pages
    Satisfactory Essays
  • Satisfactory Essays

    For ‘EXPERIMENTAL’ tubes – state the purpose, results AND interpret the results of the experiment.…

    • 592 Words
    • 3 Pages
    Satisfactory Essays
  • Powerful Essays

    Unit two Biology

    • 7492 Words
    • 30 Pages

    As the two tubes are connected by the manometer these pressure changes are cancelled out. Pressure changes affecting one tube but not the other will cause the manometer fluid to show a difference in height on the two sides of the U-tube. The apparatus is assembled with both three-way taps open to the air, to prevent the manometer fluid being pushed into the tubes. However, during an experiment both three-way taps are closed to prevent air movement between the air in the apparatus and air in the laboratory, as shown in Figure 2.18. The difference between the levels of manometer fluid on the two sides of the U-tube represents the pressure difference between the air in the two tubes.…

    • 7492 Words
    • 30 Pages
    Powerful Essays
  • Powerful Essays

    Nt1310 Unit 6 Lab Conclusion

    • 2742 Words
    • 11 Pages

    The mass of a cylinder was determined using a digital balance. These results are more questionable since the device may not have been properly calibrated. A more refined approach would be to use a newly calibrated balance of even greater…

    • 2742 Words
    • 11 Pages
    Powerful Essays
  • Good Essays

    Sound Wave

    • 564 Words
    • 3 Pages

    open end of a tube, a wave compression passes along the tube. When it reaches…

    • 564 Words
    • 3 Pages
    Good Essays
  • Good Essays

    The fluid in each of the test tubes will be analysed using a colorimeter and compared against the control, which is distilled water to check for any variations in the colour of the water.…

    • 1207 Words
    • 5 Pages
    Good Essays
  • Powerful Essays

    4. By releasing the tap gradually or pressing the pump harder, the reading of the…

    • 2729 Words
    • 11 Pages
    Powerful Essays
  • Better Essays

    In this experiment, The purpose of this experiment is to investigate the measurement of the actual volume contents of volumetric glassware. In the beginning of the experiment, the volumetric glassware should be clean and dry before used. The volumetric glassware, measuring cylinder and pipette should be handled with care and all the precautions were be taken during the experiment was held. This was to ensure to avoid any errors such as parallax error especially while reading water meniscus. This experiment must be repeated three times or more and take the average reading to get more accurate data. By using the formulae, the standard deviation and relative standard deviation could be calculated. As the result, the lower the standard deviation, the higher the consistency of an instrument. Although the instrument may measure a given sample, the value might need to be altered due to various conditions and errors.…

    • 1654 Words
    • 48 Pages
    Better Essays
  • Powerful Essays

    venturi meter

    • 914 Words
    • 4 Pages

    apparatus itself. Try and set up the apparatus initially so that the variation in pressure between the…

    • 914 Words
    • 4 Pages
    Powerful Essays
  • Good Essays

    Physics Experiment

    • 5453 Words
    • 22 Pages

    3. Mark the position of extreme node leaving out the first and last loop. Measure the length of the vibrating cord and divide it by the no. of loops to get ‘l’ for some tension.…

    • 5453 Words
    • 22 Pages
    Good Essays
  • Powerful Essays

    SuppEx Solution 4A E

    • 16750 Words
    • 145 Pages

    reaction can be followed by measuring the pressure inside the vessel with a pressure sensor connected…

    • 16750 Words
    • 145 Pages
    Powerful Essays
  • Good Essays

    The Fist

    • 469 Words
    • 2 Pages

    18. The full-wave rectified sine wave shown in below has a delay angle of 60°. Calculate Vave and Vrms in terms of Vm…

    • 469 Words
    • 2 Pages
    Good Essays